Calibrating a water level sensor is not a single procedure that applies to every device.

A float switch may only require verification of its switching height. An adjustable capacitive point sensor may use an empty- or full-liquid teach function. A hydrostatic transmitter may require zero, span and density settings, while an ultrasonic or radar sensor normally needs the empty distance, full distance and application parameters configured.

Before making any adjustment, identify the sensor technology, output type and model-specific calibration method. Applying the wrong procedure can make an otherwise functional sensor less reliable.

This guide explains how to calibrate the main types of water level sensors, how to verify the result and how XKC automatic sensitivity calibration works on a compatible non-contact point sensor.

Key Takeaways

  • Not every water level sensor has a field-adjustable calibration function.

  • Calibration, adjustment, configuration and functional testing are related but different tasks.

  • Point-level switches are calibrated or taught at a switching position; they do not measure the entire tank range.

  • Continuous transmitters normally require at least two known reference points.

  • Record the as-found readings before changing zero, span or sensitivity.

  • Use the actual container, liquid and mounting position when calibrating a through-wall capacitive sensor.

  • Do not use PLC scaling or software offsets to conceal an installation, wiring or sensor problem.

  • Verification should include rising and falling levels, not only one static condition.

  • Calibration acceptance limits should come from the sensor specification and process requirement, not a universal percentage.

  • XKC calibration steps vary by model; a procedure for the XKC-Y29A should not automatically be applied to another XKC sensor.

Calibration, Adjustment, Configuration and Verification

These terms are often used interchangeably, but they describe different actions.

Action What It Means Example
Calibration Comparing the sensor with a known reference and determining its error Comparing a transmitter reading with a measured water depth
Adjustment Changing the sensor to reduce the identified error Correcting zero or span
Configuration Entering application parameters Setting tank height, density, damping or output range
Teach or sensitivity setting Allowing a point sensor to learn an empty or liquid-present condition Teaching a capacitive sensor through a plastic wall
Verification Confirming that the completed system performs within the required limits Checking several levels after adjustment

In everyday maintenance language, “calibration” often includes comparison, adjustment and verification. A controlled procedure should still record these stages separately.

This distinction matters because repeated adjustment without first measuring the error removes useful evidence about sensor drift.

Identify the Sensor Before Calibrating It

The correct procedure depends on what the sensor measures and how it produces an output.

Sensor Type Normal Field Procedure Required Reference Important Limitation
Mechanical float switch Verify or reposition the switching point Measured liquid height Many float switches have no electronic calibration
Conductive probe Verify probe height and controller response; adjust controller sensitivity only when supported Known liquid position Deposits between probes can imitate a liquid signal
External capacitive point sensor Teach or adjust sensitivity in dry and liquid-present conditions Actual wall, liquid and switching position One sensor provides one level point
Hydrostatic level transmitter Verify pressure or level, then adjust zero and span if required Known depth or reference pressure Liquid density and vessel pressure affect the result
Ultrasonic transmitter Configure empty and full distances and verify readings Measured distance or level Foam, condensation and obstructions may affect echoes
Radar transmitter Configure measurement range and application parameters Known tank geometry or level Incorrect nozzle position or false echoes can affect measurement
Continuous capacitive transmitter Set empty and full reference values Known minimum and maximum levels Liquid properties and probe buildup can change response
RS485 or other digital sensor Compare the transmitted engineering value with a reference Known level or distance Communication configuration is not sensor calibration

First determine whether the device is:

  • A point-level switch

  • A multi-point level system

  • A continuous level transmitter

A point-level sensor only determines whether liquid has reached its installed position. It does not have a complete empty-to-full span unless it is part of a larger multi-sensor system.

When Should a Water Level Sensor Be Calibrated?

Calibration or verification may be appropriate:

  • During initial commissioning

  • After the sensor has been relocated

  • After the container or pipe has been replaced

  • When the wall material or thickness changes

  • When a different liquid will be detected

  • After cleaning or mechanical maintenance that affects the sensor

  • After replacing a transmitter, controller or input module

  • When readings disagree with a reliable physical reference

  • After an event that may have damaged or shifted the sensor

  • At an interval established by the maintenance program

Calibration should not be the first response to every incorrect reading. A sensor may appear inaccurate because of:

  • Incorrect supply voltage

  • Reversed or damaged wiring

  • Incompatible NPN, PNP, voltage or current inputs

  • A loose mounting surface

  • An air gap beneath an external sensor

  • Deposits on an immersed probe

  • A blocked pressure reference tube

  • Foam, turbulence or condensation

  • Incorrect tank dimensions in the controller

  • Faulty PLC scaling

Correct these conditions before changing calibration values.

Prepare for Calibration

1. Confirm the Complete Model Number

Obtain the sensor manual and confirm:

  • Supply voltage

  • Output type

  • Measuring range

  • Adjustment interface

  • Permitted container material

  • Required reference conditions

  • Indicator behavior

  • Calibration timing

  • Factory default settings

The same product family may contain voltage, NPN, PNP, analog and communication versions. Their calibration and verification methods may differ.

2. Isolate Controlled Equipment

Prevent the calibration process from unexpectedly starting:

  • Pumps

  • Valves

  • Heaters

  • Chemical dosing equipment

  • Overflow alarms

  • Automatic refill systems

The sensor and controller may remain powered when required by the calibration procedure, but connected actuators should be placed in a safe commissioning state.

3. Inspect the Installation

Before adjusting the sensor:

  • Clean the sensing area according to the manufacturer's instructions.

  • Check that the sensor has not moved.

  • Confirm that an external sensor is firmly attached.

  • Inspect the cable and connector.

  • Check the power supply at the sensor.

  • Verify that the liquid is reasonably stable.

  • Confirm that the selected reference point can be measured safely.

Do not electronically compensate for a loose sensor, a blocked pressure port or heavy buildup.

4. Select a Reference

Possible references include:

  • A calibrated dip tape

  • A sight glass

  • A measured fill volume in a tank with known geometry

  • A reference pressure source

  • A certified distance instrument

  • A known liquid surface position

  • A calibrated simulator for checking the output loop

The reference must be more reliable than the sensor being evaluated.

5. Define the Acceptance Limit

Decide how much error is acceptable before starting.

The limit should consider:

  • Manufacturer accuracy specification

  • Required switching height

  • Process control requirement

  • Tank dimensions

  • Overflow or dry-run consequences

  • Repeatability requirement

  • Applicable site quality procedures

There is no universal error percentage that determines whether every water level sensor passes or fails.

6. Record the As-Found Condition

Before changing anything, record:

  • Actual reference level

  • Sensor reading or output

  • Display or PLC reading

  • Alarm state

  • Rising or falling direction

  • Existing zero, span or sensitivity setting

  • Liquid temperature when relevant

  • Visible installation conditions

These records show whether the sensor was actually out of tolerance and how much adjustment was required.

General Water Level Sensor Calibration Procedure

A controlled calibration normally follows this sequence.

Step 1: Stabilize the Process

Allow turbulence, foam and vibration to settle as much as practical. For temperature-sensitive applications, allow the sensor and liquid to reach representative operating conditions.

Step 2: Check the Lower Reference

Place the liquid at the required empty, low-level or zero reference. Record the physical reference and sensor output.

Do not automatically assume that an empty tank should equal zero. Some installations use an elevated or suppressed zero because the sensor is mounted below the tank bottom or because a pressure transmitter includes additional static head.

Step 3: Check the Upper Reference

Raise the liquid to the required high-level or full-scale reference. Allow the reading to stabilize and record the result.

The upper calibrated range value is the maximum level the system is configured to measure. It is not necessarily the sensor's absolute mechanical or electrical limit.

Step 4: Check Intermediate Points

For a continuous transmitter, check intermediate values when practical. A five-point verification might use:

  • 0%

  • 25%

  • 50%

  • 75%

  • 100%

This can reveal a linearity problem that would remain hidden during a two-point check.

Step 5: Check Rising and Falling Levels

Repeat selected points while lowering the liquid. Comparing rising and falling readings reveals hysteresis, mechanical sticking and reset-point behavior.

Step 6: Adjust Only When Necessary

If the error exceeds the defined acceptance limit, use the manufacturer's procedure to adjust:

  • Zero

  • Span

  • Sensitivity

  • Empty distance

  • Full distance

  • Density

  • Offset

  • Output trim

Avoid changing several parameters at once. Otherwise, it becomes difficult to determine which adjustment corrected or introduced the error.

Step 7: Perform an As-Left Verification

After adjustment, repeat the required reference points and document the final readings. A successful adjustment is not complete until its result has been independently verified.

How to Calibrate a Point-Level Sensor

A point-level sensor switches when liquid reaches one position.

Its important calibration characteristics are:

  • Switching height

  • Reset height

  • Repeatability

  • Normal output state

  • Response delay

  • Sensitivity to residue or external moisture

Float and Mechanical Level Switches

Many mechanical float switches do not have zero and span controls.

To verify one:

  1. Inspect the float for free movement.

  2. Raise the liquid slowly.

  3. Measure the height at which the contact changes state.

  4. Continue above the switching point.

  5. Lower the liquid slowly.

  6. Measure the reset height.

  7. Repeat the cycle several times.

If the device has adjustable collars, counterweights or tether lengths, reposition them according to the manual. Do not bend a float arm or modify a sealed mechanism unless the manufacturer specifically permits it.

Conductive Level Probes

Before changing controller sensitivity:

  1. Clean the probes and surrounding insulation.

  2. Confirm the probe heights.

  3. Check for residue bridging adjacent probes.

  4. Test with the actual liquid.

  5. Adjust the controller only within its documented range.

  6. Verify both filling and emptying conditions.

Excess sensitivity may cause a conductive film or foam to be interpreted as a full liquid level. Insufficient sensitivity may prevent low-conductivity liquids from being detected.

External Capacitive Point Sensors

A through-wall capacitive sensor should normally be calibrated in its final installation because its response depends on the complete sensing path:

  • Liquid

  • Container material

  • Wall thickness

  • Mounting pressure

  • Air gaps

  • Nearby metal

  • Residue

  • External moisture

Calibrating the sensor in your hand and then attaching it to the tank does not reproduce the final electrical environment.

How to Calibrate an XKC Non-Contact Water Level Sensor

XKC manufactures several non-contact liquid level sensor models, but their calibration interfaces are not identical. Always use the manual for the complete product version.

The XKC-Y29A wall-mounted liquid sensor provides a useful example because it supports automatic sensitivity calibration through an onboard button or the black MARK calibration wire.

The XKC-Y29A is a point-level sensor for compatible non-metallic containers. Calibration teaches the sensor to distinguish liquid-present and liquid-absent conditions through the container wall. It does not create a continuous 0–100% tank measurement.

Prepare the XKC-Y29A

Before calibration:

  1. Confirm that the product is an XKC-Y29A and identify its output version.

  2. Mount the sensor at its final level position.

  3. Use a compatible non-metallic container.

  4. Keep the sensing face firmly against the wall.

  5. Eliminate any unnecessary air gap.

  6. Avoid metal brackets, tank ribs and joints around the sensing area.

  7. Use the actual liquid that will be present in operation.

  8. Power and wire the sensor according to its model-specific manual.

  9. Isolate any pump or valve controlled by the sensor.

The current Y29A documentation identifies:

  • Brown as VCC

  • Blue as GND

  • Yellow as output

  • Black as the MARK calibration wire

These definitions must not be assumed for another XKC model.

Full-Liquid Calibration

Full-liquid calibration teaches the sensor using a liquid-present condition at the required switching position.

For the current XKC-Y29A procedure:

  1. Raise the liquid until the surface is approximately level with the center of the installed sensor.

  2. Keep the liquid stable.

  3. Press the calibration button, or momentarily operate the MARK-to-GND calibration connection exactly as shown in the manual.

  4. Hold it for approximately one second and then release it.

  5. Wait while the indicator flashes.

  6. Do not interrupt power until the calibration sequence finishes.

  7. Confirm the final indicator and output state.

The full-liquid teach establishes a switching threshold below the detected liquid-present reference. The exact timing and indicator sequence should always be checked against the manual revision supplied with the product.

Empty-Liquid Calibration

Empty-liquid calibration uses a condition in which liquid is absent from the sensing position.

For the current XKC-Y29A procedure:

  1. Lower the liquid completely below the sensor.

  2. Confirm that the intended “empty” condition is present at the sensing area.

  3. Press the calibration button or operate the prescribed MARK-wire connection.

  4. Hold it for approximately five to six seconds.

  5. Release it when required by the manual.

  6. Wait for the calibration indication to finish.

  7. Confirm the resulting dry-state output.

An empty teach may be useful when the real empty condition includes a normal residual film, moisture or foam. However, it should not be used to conceal an incompatible wall, a large air gap or a mounting position affected by external water.

MARK Wire Precautions

The MARK wire is a calibration input, not a normal sensor output.

  • Use it only as described in the applicable manual.

  • Do not leave it permanently connected unless the wiring diagram specifically requires that state.

  • Do not apply an arbitrary external voltage.

  • Insulate it when the manual requires it to remain open.

  • Do not assume another XKC model uses the same hold time.

Some XKC sensors use different calibration durations, indicators or logic. Reusing the Y29A sequence on another model could start the wrong teach mode or fail to save the intended reference.

Verify the XKC Calibration

After calibration:

  1. Lower the liquid below the sensing position.

  2. Record the indicator and output.

  3. Raise the liquid slowly through the sensor position.

  4. Record the switching height.

  5. Raise the liquid above the sensor.

  6. Lower it through the switching position.

  7. Record the reset height.

  8. Repeat the complete cycle at least three times.

  9. Confirm that the receiving controller interprets both states correctly.

  10. Repeat the test under representative residue, temperature and moisture conditions when these factors are important.

If the sensor changes state correctly but the controller does not, the problem is more likely to involve wiring, input compatibility or controller logic than sensor sensitivity. The water level sensor testing guide explains how to distinguish these faults without repeating the calibration process.

How to Calibrate a 4–20 mA Level Transmitter

A 4–20 mA transmitter normally maps a configured lower range value to 4 mA and an upper range value to 20 mA.

For a direct-acting linear output:

Output current (mA) = 4 + 16 × [(Reference Level − LRV) ÷ (URV − LRV)]

Where:

  • LRV is the lower range value.

  • URV is the upper range value.

  • The reference level is the actual measured level.

For example, if 0 m represents 4 mA and 2 m represents 20 mA, a verified level of 1 m should nominally correspond to 12 mA.

This calculation only applies when the transmitter is configured for a direct, linear level output. Reverse outputs and systems with tank-volume linearization require their configured transfer function.

Calibration Procedure

  1. Confirm the LRV and URV.

  2. Verify that the analog input is configured for current.

  3. Measure the actual level with a suitable reference.

  4. Measure the loop current with an appropriate instrument.

  5. Compare the measured current with the expected current.

  6. Repeat at several reference levels.

  7. Perform zero or span adjustment only if the sensor itself is outside tolerance.

  8. Check the 4–20 mA output separately with a loop test when supported.

  9. Repeat the complete verification after adjustment.

A sensor trim and an output trim are not the same operation. A sensor trim corrects the relationship between the physical input and the transmitter's digital measurement. An output trim corrects the relationship between the transmitter's digital value and its analog current output.

Do not adjust the 4 mA and 20 mA outputs to compensate for an incorrect level measurement without first identifying which part of the measurement chain is wrong.

How to Calibrate a 0–10 V Level Sensor

For a direct linear output configured from 0 to 10 V:

Expected output voltage (V) = 10 × [(Measured Value − LRV) ÷ (URV − LRV)]

The procedure is similar to a current-output transmitter:

  1. Confirm the configured level range.

  2. Verify the signal reference and analog-input mode.

  3. Measure the actual level.

  4. Measure the output voltage.

  5. Compare it with the expected value.

  6. Check several rising and falling levels.

  7. Adjust zero or span only according to the model manual.

Voltage errors may also result from ground-potential differences, cable resistance, electrical noise or an incorrect signal reference. Recalibrating the sensor will not correct these circuit problems.

How to Calibrate a Hydrostatic Level Sensor

A hydrostatic sensor measures the pressure produced by the liquid column above its sensing diaphragm.

Calibration therefore depends on:

  • Liquid height

  • Liquid density

  • Atmospheric or vessel pressure

  • Sensor mounting elevation

  • Venting arrangement

  • Configured pressure range

Basic Procedure

  1. Confirm the required liquid density or specific gravity.

  2. Inspect the pressure port and vent tube.

  3. Establish a verified low-level reference.

  4. Record the sensor output.

  5. Establish one or more higher reference levels.

  6. Allow the pressure to stabilize at each point.

  7. Compare the measured level and analog output with the reference.

  8. Adjust the zero, span or density parameter only as specified.

  9. Verify the result while lowering the level.

Do not assume that exposing a submersible sensor to air always represents the correct installed zero. Mounting elevation, sealed-tank pressure and configured range suppression may make the correct zero different from atmospheric pressure at the loose sensor.

A blocked vent tube, damaged diaphragm or accumulated material must be corrected physically. Electronic zero adjustment should not be used to hide these conditions.

How to Calibrate an Ultrasonic or Radar Level Sensor

Ultrasonic and radar transmitters determine level from the distance between the sensor and the liquid surface.

Their basic setup normally uses:

  • Empty distance

  • Full distance or span

  • Sensor reference point

  • Blocking or dead distance

  • Tank geometry

  • Application and echo parameters

Basic Procedure

  1. Measure from the sensor's defined reference point to the empty-level reference.

  2. Enter the empty distance according to the manual.

  3. Measure the distance to the maximum required level.

  4. Enter the full distance or measuring span.

  5. Confirm that the full point remains outside the prohibited blocking distance.

  6. Review false echoes from nozzles, supports, ladders or internal equipment.

  7. Compare the displayed level with a physical reference at several points.

  8. Check the analog or digital output separately.

  9. Save and document the final configuration.

Some radar sensors can be configured from known vessel dimensions without filling and emptying the tank. This is still followed by verification under actual process conditions.

For ultrasonic sensors, foam, heavy vapor, condensation, temperature gradients and an uneven surface can affect the echo. Increasing software damping may stabilize a fluctuating display, but it does not correct an incorrect measurement range or poor mounting position.

Do Not Confuse Level Calibration with Tank-Volume Calibration

A level transmitter measures height or distance. A controller may then convert that level into volume.

For a vertical rectangular tank, level and volume may have a linear relationship. For other vessels, they may not.

Examples include:

  • Horizontal cylindrical tanks

  • Conical-bottom tanks

  • Spherical tanks

  • Irregular mobile tanks

  • Tanks with internal displacement

  • Vessels with multiple sections

If the level value is correct but the displayed volume is wrong, the sensor may not require calibration. The controller may instead need a corrected tank geometry, strapping table or linearization curve.

Calibration Acceptance Checklist

Point-Level Sensor

Confirm:

  • Correct dry-state output

  • Correct liquid-present output

  • Switching at the required height

  • Reset at an acceptable height

  • Stable operation through repeated cycles

  • No false triggering from normal residue or exterior moisture

  • Correct interpretation by the receiving controller

Continuous Level Transmitter

Confirm:

  • Error at each required reference point

  • Rising and falling readings

  • Repeatability

  • Zero and span

  • Analog or digital output

  • Local display and controller agreement

  • Alarm activation points

  • Valid behavior outside the normal range

  • Correct tank-volume conversion when used

A single correct reading does not demonstrate accuracy across the entire range.

Why Calibration May Fail

Symptom Likely Cause Calibration Response
No stable empty or full state Turbulence, foam, residue or loose mounting Stabilize or correct the installation before teaching
External capacitive sensor always detects liquid Excess sensitivity, inner-wall film, external water or nearby metal Inspect conditions before performing an empty teach
Sensor works before mounting but not on the tank Wall incompatibility, excessive thickness or an air gap Correct the application rather than repeatedly recalibrating
Sensor output is correct but PLC value is wrong Incorrect input type or scaling Correct the controller configuration
Analog output is stable but offset at every point Zero or installation offset Verify the reference before performing a zero adjustment
Error increases toward the top of the range Incorrect span, density or range configuration Check the upper reference and span
Rising and falling values differ substantially Mechanical friction, buildup or sensor hysteresis Inspect the sensing mechanism and compare with specifications
Calibration changes after power cycling Teach sequence not completed or settings not stored Repeat the model-specific saving procedure
New liquid is detected differently Different dielectric constant, density or acoustic behavior Recalibrate only if the sensor is suitable for the new medium
Frequent recalibration is required Process buildup, mechanical movement, electrical instability or sensor deterioration Correct the root cause and evaluate sensor condition

How Often Should a Water Level Sensor Be Calibrated?

There is no universal annual, quarterly or monthly interval for every sensor.

The appropriate interval depends on:

  • Manufacturer recommendations

  • Consequences of an incorrect reading

  • Required accuracy

  • Process stability

  • Liquid properties

  • Buildup or corrosion

  • Temperature and pressure cycling

  • Historical calibration drift

  • Quality-system requirements

  • Whether independent overflow or dry-run protection is available

A new installation may require more frequent checks until enough history is available. If repeated records show little drift, the maintenance program may support a different interval. Any change should follow the site's documented risk and quality procedures.

Immediate verification may be appropriate after:

  • Sensor replacement

  • Tank modification

  • Process-liquid change

  • Major cleaning

  • Mechanical impact

  • Wiring repair

  • Controller replacement

  • An unexplained overflow, dry-run event or false alarm

What Should a Calibration Record Include?

A useful record should contain:

  • Sensor manufacturer and model

  • Serial number or asset number

  • Installation location

  • Sensor technology

  • Output type and range

  • Liquid and container details

  • Reference instrument

  • Reference instrument calibration status

  • Ambient and liquid conditions

  • As-found readings

  • Adjustments made

  • As-left readings

  • Rising and falling results

  • Acceptance limits

  • Pass or fail decision

  • Calibration date

  • Technician

  • Recommended next verification date

  • Notes about residue, damage or installation changes

Trend the as-found error over time. This provides more useful evidence about sensor stability than repeatedly resetting the device without retaining its previous condition.

Conclusion

Calibrating a water level sensor begins with identifying what the device actually measures. A point-level switch, external capacitive sensor, hydrostatic transmitter and radar instrument require different reference conditions and adjustment procedures.

For an XKC teach-capable non-contact sensor such as the XKC-Y29A, sensitivity should be established on the actual non-metallic container with the final liquid and mounting position. Full- and empty-liquid teach functions set the point-detection threshold; they do not convert the sensor into a continuous level transmitter.

For continuous sensors, record the as-found condition, compare several known reference points, adjust only the parameter responsible for the error and complete an as-left verification. When readings remain unstable, correct installation, buildup, wiring or process conditions rather than repeatedly recalibrating the device.

A documented, technology-specific procedure provides more reliable level control than applying the same empty-and-full adjustment to every water level sensor.

FAQ

Can Every Water Level Sensor Be Calibrated?

No. Many float switches and fixed probes have no field calibration control. They can be tested and repositioned when their design permits it. Adjustable capacitive sensors, continuous transmitters and intelligent instruments may provide sensitivity, zero, span or teach functions.

Is Calibration the Same as Testing a Water Level Sensor?

No. Testing confirms whether the sensor and output work. Calibration compares the measurement with a known reference and determines the error. Adjustment may then be performed if the error is outside the permitted limit.

Can I Calibrate a Sensor Using Only a Multimeter?

A multimeter can measure continuity, voltage or current, but it does not establish the actual liquid level. You also need a reliable physical reference such as a known depth, distance, volume or pressure.

Can I Calibrate a Continuous Sensor with an Empty Tank Only?

An empty reference may establish zero, but it does not verify span or linearity. Continuous measurement should normally be checked at additional known points or configured using validated tank dimensions and then verified under operating conditions.

Does an XKC Non-Contact Sensor Always Need Calibration?

It depends on the model and installation. A factory setting may work in a typical application, but a teach-capable model may require sensitivity calibration when the liquid, container material, wall thickness or mounting environment differs from the original reference conditions.

Can One XKC Point Sensor Be Calibrated to Show Tank Percentage?

No. A point sensor only indicates whether liquid is present at its installed height. Several sensors can provide multiple level thresholds, but continuous percentage measurement requires a suitable continuous sensor or sensor array.

Should I Calibrate with Water Before Using Another Liquid?

Use the actual process liquid whenever practical. A capacitive sensor may respond differently to oil, detergent, syrup or another liquid. Hydrostatic measurement also depends on density, while ultrasonic performance can be affected by foam or vapor conditions.

Why Did the Sensor Become Less Reliable After Calibration?

The wrong state may have been taught, the liquid may have been unstable, or the sensor may have been calibrated before final installation. Incorrect hold time, an air gap, external moisture, residue or use of another model's procedure can also produce a poor result.

Should I Adjust the PLC When the Sensor Reading Is Wrong?

First compare the sensor output with a physical reference. If the sensor output is correct but the PLC value is wrong, correct the PLC input configuration or scaling. If the sensor output itself is wrong, investigate the sensor and installation before changing software.

How Do I Know Whether Calibration Was Successful?

The sensor should meet the predefined acceptance limit at the required reference points, operate repeatably during rising and falling levels and produce the correct output at the receiving controller. The final result should be documented as an as-left verification.